1. Illuminance Meter
      2. model no. : T-10A / 10MA


      3. Desc. :

      4. KONICA MINOLTA



      5. Application :

      6. Indoor lighting, sun light, light box for advertisement, street light, tunnel lighting, ANSI lumen of projector, effect of light to animals or plants



      7. Feature :

      8. To measure light intensity with lux or fcd, wide range, accurate, easy to use, compact, data output to PC, support multi-point measurement



  • Model: T-10A (standard receptor); T-10MA (mini receptor)
    Standard compliance: JIS C 1609-1:2006 class AA
    Sensor: Silicon photocell
    Relative spectral response: Within 6% (f1´) of the CIE spectral luminous efficiency V (λ)
    Cosine Correction Characteristics:

    Within ±1% at 10°; Within ±2% at 30°; Within ±6% at 50°;

    Within ±7% at 60°; Within ±25% at 80°

    Illuminance unit: lux (lx) or foot candles (fcd) (switchable)
    Measurement range:
    Illuminance : 0.01 ~ 299,900 lx, 0.001 ~ 29,990 fcd
    Integrated illuminance : 0.01 ~ 999,900 x klx•h,
    0.001 ~ 99,990 x kfcd•h / 0.001 ~ 9999 h
    User calibration function: CCF (Color Correction Factor) setting function
    Accuracy:

    Within ±2% ±1 digit of displayed value

    (based on Konica Minolta standard)

    Temperature / humidity drift:

    Within ±3% ±1 digit of displayed value

    (at 20°C / 68°F within operating temperature / humidity range)

    Digital output: USB
    Display: 3 or 4 significant digits (LCD with back-light illumination < 10 lx)
    Power source: 2 AA-size batteries or AC adapter (optional)
    Dimensions:

    Main body & standard receptor : 69 x 174 x 35 mm;

    main body : 69 x 161.5 x 30 mm 
    mini receptor : Ø16.5 x 13.8 mm; cord length : 1 m

    Weight: 200 g (without battery); 205 g (without battery)